Valleylab Force 2 Electrosurgical Generator w/ monopolar foot switch
OPTIONS:
- Bipolar foot switch and
- Rolling cart.
Standard Accessories Available
- Handpieces
- Grounding Pads
Features:
- Provides cutting and coagulation in monopolar or bipolar outputs
- Standard REM (Return Electrode Monitoring)
- Has both audible and visual alarms, including REM™ Fault Indicator.
- High Power Digital ESU
- High performance capabilities in a multipurpose generator
- Provides flexibility through varying degrees of homeostasis
- Operator can choose one of three preset modes: 1)simultaneous independent coagulation, 2) low voltage coagulation and 3) Valleylab REM™ safety
- Varying degrees of homeostasis.
- Simultaneous independent coagulation.
- Low voltage coagulation.
- Valleylab REM safety.
- Reduces the risk of patient burns.
- Weight 23lbs
- 300 watts is available for output in the cut setting.
BLENDED CUT MODES
Provide flexibility through varying degrees of hemostasis. Choose one of three preset modes.
SIMULTANEOUS INDEPENDENT COAGULATION
Permits two surgeons to fulgurate from a single generator, for added convenience and efficacy. Accessories are activated only when keyed, reducing the potential for injury caused by inadvertent activation.
LOW VOLTAGE COAGULATION
Ensures controlled, precise desiccation with less destruction of peripheral tissue, making the Forceª 2 generator ideal for laparoscopic procedures.
REM™ SAFETY
Is guaranteed with Valleylab's patented adaptive REMª system, which continually monitors patient impedance levels. If a fault in the patient/return electrode contact is detected, the REMª system automatically deactivates the generator virtually eliminating the risk of burns under the return electrode. Valleylab REMª safety has been proven in more than 100 million surgical procedures worldwide.
WEIGHT AND DIMENSIONS
- Height: 8 in. (20 cm)
- Width: 13 in. (33 cm)
- Length: 21 in. (53 cm)
- Weight: 23 lbs. (10.4 kg)
OUTPUT WAVEFORMS LOW FREQUENCY LEAKAGE (50-60 Hz) HIGH FREQUENCY LEAKAGE INPUT POWER REQUIREMENTS POWER READOUTS ADAPTIVE REM™ SYSTEM COOLING AUDIO VOLUME
Source current, patient leads, all outputs tied together.
Less than 150 mA rms
Operating range is 85 to 135 AC volts. Current is less than 8 amperes in cut and less than 4 amperes in coag.
Agree with actual power into rated load to within ± 15% or 5 watts, whichever is greater.
--REMª pad Ñ 5-135 ohms
--Non-REMª pad Ñ less than 20 ohms
Convection, no fan
The mode indicator tones are adjustable to a minimum level of 45 dB at 1 meter. The alarm tones are not adjustable and are set at 65 dB minimum at 1 meter.
Also available in a 220-240 volt, 50-60 Hz configuration. Designed to meet UL and CSA specifications.
The Force 2 electrosurgical unit from Valleylab provides cutting and coagulation in monopolar or bipolar outputs. Force 2s have audible and visual alarms, including REM™ Fault Indicator. The REM™ (Return Electrode Monitoring) is standard. 300 watts is available for output in the cut setting. A monopolar foot switch is included, a bipolar foot switch is optional as is a rolling cart.