Lighthead provides cool, shadow-reduced, color-corrected light appropriate for surgical procedures. For optimal color rendition in cavity illumination.
- Lighthead design prevents direct glare and disposes of 98% of its generated heat
- Tungsten-halogen lamp emits energy which is filtered and reflected by two dichroic, coated, glass surfaces.
- Two aluminum reflectors pass the controlled visible light beam to the surgical site.
- Optical lens diffuses and focuses the light beams onto the surigcal field.
- Vinyl bumper protects the outer cover and large reflector.
- Horizontal Suspension Arm Assembly supports a movable yoke into which the lighthead is mounted
Specifications
- Maneuverable: With the lighthead in the horizontal plane it shall be possible to tilt the reflector 320 degrees forward and backward 540 degrees laterally, left an right.
- Light Color and Shadows: Light projection exceeds recommendations for shadow reduction according to IES Guide CP29 LIghting for Health Care Facilities.
- Focus / Dept of Field: Standrd patterns and intensities are obtained at a distance of 38†from the surgical field. Depth of field is a minimum of 29†for all pattern sizes.
- Beam Temperature: The radiant heat energy in the light beam, 38†(965mm) below the reflector rim, shall not exceed 13,000 microwatts per cm, when measured by a thermopile (based upon readings taken with the lighthead at 4000fc intensity and after the temperature has stablized following continuous for at least one hour.